- The portable MAS-100 Eco® sampler is the basic version of the MAS-100® range. It is used for microbiological air monitoring in the hygienic production of food and cosmetics, as well as in the maintenance of ventilation and air-conditioning systems (HVAC). The MAS-100 Eco® sampler is the lightest of the models and particularly easy to handle.
- Flow rate: 100 liters per minute (LPM) ±4%
- Compatible with different nutrient plates, e.g., ready-to-use 90 mm standard plates, and (with optional adapters) 55 mm contact or Growth Direct plates
- Compliant with ISO 14698-1 and ISO 14698-2
Material: Body / Perforated lid: Anodized aluminum / anodized aluminum
Flow rate: 100 liters per minute (LPM) ±4%
Sample volumes: 1 to 1000 liters (pre-set: 10, 20, 100, 250 and 500 L)
Compatible with different nutrient plates, e.g., ready-to-use 90 mm standard plates, and (with optional adapters) 55 mm contact or Growth Direct plates
Autoclaving (wet or dry)
Temperature: ≤ 180 °C
Duration: < 60 min"
Rechargeable Batteries: 2 NiMH rechargeable batteries, 1.2 V
Battery capacity: Approx. 3 h (= 18.000 L)
Blower motor: 6 Vdc
Display: Alphanumeric liquid crystal display, 2 x 8 characters
Lifetime RTC battery: Approx. 10 years
Dimensions of perforated lid (D × H): Ø 109 × 32 mm
Dimensions of body (H × D) 190 × 110 mm (with handle)
Weight 1.4 kg
Sample modes: Standard single sample, start delay configurable from 1 to 60 min
Power supply input: 100 - 240 V AC, 50 - 60 Hz, 0.4 A to 1.0A (depending on power supply model)
Temperature Storage: Operation: 5 to +40 °C
Maximum relative humidity: Max. 80 % RH for temperatures up to 31 °C, decreasing linearly to 50 % relative humidity at 40 °C; noncondensing