Critical applications
• Microbiology culture media preparation
• Staining solutions for histology and cytology
• Immunohistochemistry (IHC)
• Gel electrophoresis, western blotting
• Immunoassays (ELISA, RIA)
• Dissolution testing
• Biological Oxygen Demand (BOD)/ Chemical
Oxygen Demand (COD)
• UV/Vis Spectroscopy
• Titration"
General laboratory applications
• Sample preparation (dilution, extraction, etc.)
• Buffer and reagent preparation
• Glassware rinsing"
Equipment & instruments
• Autoclaves
• Dishwashers
• Weathering and stability test chambers
• Clinical analyzers and slide stainers
• Hydrogen generators
• Ultrapure water systems (e.g. Milli-Q® IQ 7000 system)"
With respect to daily, freshly produced water, the Milli-Q® IX system is intended to dispense pure water that meets or exceeds water quality specifications described by the organizations below: European Pharmacopoeia - Purified water, U.S. Pharmacopeia - Purified water, Japanese Pharmacopoeia - Purified water, Chinese Pharmacopoeia - Purified water, ASTM® D1193 - Type 2 water - Reagent water, ISO® 3696 - Grade 2 water, Chinese National Standard - GB 6682 Grade 2 water
Set up convenient water stations throughout your lab by connecting two E-POD® dispensers up to 5 m apart.
Keep working as Volumetric Dispense mode fills glassware to your set volume (1 ml increments from 20 mL to 99 L). Simply push the button, the screen or the foot pedal to repeat
Interact with ease. The smartphone-like touchscreen allows intuitive use, even with gloves on!
Capacitive touchscreen Size: 5 in; Resolution: 800 x 480"
Complementary purification technologies remove contaminants to deliver consistently high-quality pure water directly from tap water.
IPAK Gard® pretreatment pack
High efficiency removal of colloids, particles, free chlorine and minerals for improved system performance"
Advanced RO
• Removes 95–99% of contaminants including ions, particles, bacteria and organic molecules (MW >200 kDa)
• Decreases water consumption and ensures a constant flow rate
• RO recovery loop optimizes water recovery vs. standard RO systems and reduces water consumption"
Elix® EDI module
• Removes remaining ions for consistently superior quality pure water
• No maintenance, ensuring low and predictable running costs"
ech2o® bactericidal UV lamp
• Mercury-free UVC LED technology emitting at 265nm provides highest efficiency bacterial inactivation
Intelligent pure water storage tank provides multi-targeted prevention of bacterial, particulate and CO2 contamination.
Within the tank, pure water quality is preserved by 3 built-in features:
- Vent filter, redesigned for seamless integration, provides improved protection against airborne contaminants
- Automatic Sanitization Module (ASM) with an integrated mercury-free ech2o® UVC LED lamp emitting at 265 nm, regularly irradiates stored water and tank walls, preventing bacterial growth and biofilm formation"
Final filters adapt water quality to your needs
Automated alerts notify when consumables need replacing to avoid risk of impacting major components
All water and system data are stored in system memory, avoiding manual entries in a logbook and increasing data quality, reliability and traceability.
Rapidly access data. View data on screen, export it via the Ethernet connection or USB port on the dispenser, or email it to yourself using a report’s QR code. All dispense attributes can be quickly retrieved to prepare an audit or for QC purposes.
Feed water requirements
Feed water Potable tap water
Pressure 1–6 bar
Temperature 5–35 °C
Conductivity <2000 μS/cm at 25 °C
Free chlorine <3 ppm
Total Organic Carbon (TOC) <2 ppm
Hardness (as CaCO3) <300 ppm
Silica <30 ppm
Pure, Type 2 water specifications
Resistivity at 25 °C: >5 MΩ·cm; typically 10–15 MΩ·cm
TOC ≤30 ppb
Production flow rate: 15 L/h
Particulates: No particles with size >0.22 µm ( With Millipak® or Millipak® Gold filter)
Bacteria ≤10 cfu/L (With Millipak® or Millipak® Gold filters or Biopak® polisher when installed and used in a laminar flow hood)
Pyrogens (endotoxins) <0.001 EU/mL (With Biopak® polisher when installed and used in a laminar flow hood.)
RNases <1 pg/mL (With Biopak® polisher)
DNases <5 pg/mL (With Biopak® polisher)
Proteases <0.15 µg/mL (With Biopak® polisher)
Flow rate Up to 2 L/min
Tubing and port requirements
Feed water connection ½” Gaz
Distance from feed water port: Maximum 5 m
Distance from purification unit to E-POD® dispenser: Maximum 5 m
Distance between two POD dispensers (Maximum 2 dispensers connected in series): Maximum 5 m
Distance from purification unit to tank: Maximum 5 m
Power entry: Connection IEC 13
Ethernet port: IEEE P802.3
Electrical connections and specifications
Power source voltage 100–240 VAC ± 10%
Power frequency 50–60 Hz ± 2 Hz"
Power cord length 2.5 m (8.2 ft) plug: IEC13 female
Operational temperature 4–40 °C
Tank material High purity polyethylene
Dry weight: 22.5 kg (Milli-Q® IX 7010)
Operating weight: 27.4 kg (Milli-Q® IX 7015)
E-POD® Operating weight: 5.5 kg
Screen description and functionalities
Capacitive touchscreen Size: 5 in; Resolution: 800 x 480
USB port USB 2.0 Highspeed standard